Endress Hauser Proline Prowirl R 200 vortex flowmeter 7R2CRF 7R2CRG 7R2CRJ
The Proline Prowirl R 200 is a cost-effective vortex flowmeter from Endress+Hauser, optimized for low-flow applications and designed to deliver best-in-class accuracy even with pipe diameter reductions. Featuring integrated diameter reduction by one or two line sizes, it eliminates the need for costly pipework modifications, providing significant cost and time savings.
Key Features and Benefits
-
Designed for Low Flows and Energy Management: The Prowirl R 200 is a particularly dependable solution for energy management, dedicated to applications with very low or reduced flow. Its integrated temperature and pressure measurement enables easy energy management for steam and gases.
-
Best-in-Class Accuracy Despite Pipe Reduction: The integrated diameter reduction by one or two line sizes means no pipework modifications are needed for line size reduction. It delivers the same accuracy down to Reynolds number Re 10,000 and features one of the most linear vortex meter bodies available.
-
Long-Term Stability: The robust, drift-free capacitive DSC sensor ensures reliable measurement over extended periods without recalibration.
-
Loop-Powered Two-Wire Technology: Enables cost-effective and seamless integration into existing infrastructures while offering the highest operational safety in hazardous areas.
-
Safe and Convenient Operation: The transmitter features a separate connection compartment for convenient wiring and a 4-line backlit display with touch control for safe operation without opening the device.
-
Integrated Heartbeat Technology: Built-in verification ensures measurement reliability, enables traceable verification according to ISO 9001, and allows extension of recalibration cycles.
Technical Specifications (at a Glance)
| Specification | Detail |
|---|---|
| Measuring Principle | Vortex (Karman vortex street) |
| Nominal Diameter (Mating Pipe) | DN 25 to 250 (1 to 10″) |
| Measured Variables | Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature |
| Max. Measurement Error | Liquid volume flow: ±0.75%; Steam/gas volume flow: ±1.00%; Liquid mass flow: ±0.85% |
| Measuring Range (Liquid) | 0.1 to 540 m³/h (0.061 to 320 ft³/min) |
| Measuring Range (Steam/Gas) | 0.52 to 7,300 m³/h (0.31 to 4,300 ft³/min) |
| Max. Process Pressure | PN 40, Class 300, 20K |
| Medium Temperature Range | Standard: –40 to +260°C; Option: –200 to +400°C; On request: –200 to +450°C |
| Wetted Materials | Measuring tube: 1.4408 (CF3M); DSC sensor: 1.4404/F316/F316L |
| Transmitter Housing | AlSi10Mg, coated; 1.4404 (316L) |
| Degree of Protection | IP66/67, Type 4X enclosure |
| Display/Operation | 4-line backlit display with touch control; remote display available |
| Digital Communication | HART, PROFIBUS PA, FOUNDATION Fieldbus |
| Power Supply | DC 12 to 35 V (4-20 mA HART); DC 9 to 32 V (PROFIBUS PA) |
| Functional Safety | IEC 61508 (SIL), applicable in IEC 61511 safety applications |
| Hazardous Area Approvals | ATEX, IECEx, cCSAus, JPN, EAC, UK Ex |
| Marine Approvals | ABS, LR, BV, DNV |
The Proline Prowirl R 200 delivers reliable, cost-effective vortex flow measurement for low-flow steam, gas, and liquid applications, combining an innovative integrated diameter reduction design, robust construction, loop-powered technology, and the advanced diagnostic capabilities of Heartbeat Technology.